Patient Buddy System
Abstract
A system to enable the doctor or caregivers to remotely detect the physiologic measurements of multiple patients and receive alerts if patients’ conditions get critical or if the patient seeks emergency help. The system measures heartbeat, body temperature, position (sleeping or standing) and activity of the day (step count) continuously. The doctor or caregiver can monitor these measurements and receive alerts in their mobile app.
Objectives & Introduction
- Monitor physiologic measurements of the patients: heartbeat, temperature, patient position: sleeping or standing, patient activity: step count
- Allows the doctor/caregiver to monitor physiologic data for multiple patients.
- Determine if the patient’s physiological conditions are critical.
- Automatically alert the doctor/caregiver of the patient’s conditions if critical conditions are reached.
- Allow the patient to request for emergency help.
Team Members & Responsibilities
- Avinash Closepet Suresh
- Orientation algorithm development and implementation
- Girish Bangalore Ramesh
- Board to server communication implementation
- Thiruparan Balachandran
- Step count algorithm development, implementation and system integration
- Nimmi Bhatt
- Heart rate algorithm and implementation
- All of us:
- Collectively worked on end to end system testing and integrating data into an android app
Schedule
Week# | Date | Task | Actual |
---|---|---|---|
1 | 4/14/2015 | Order materials, ElectricImp exploration, Patient Position | Completed ordering materials |
2 | 4/21/2015 | Develop position details about the subject's orientation via the accelerometer, test different algorithm methods and store the data in Flash. | Completed getting the position details and developing a basic algorithm; Attempts to modify the algorithm are in progress. |
3 | 4/28/2015 | End to End Communication: Getting the data from the sensor and displaying it on the app's page. | Completed sending data via electric imp and reading it on the server on other side. |
4 | 5/05/2015 | Add other sensors to the system (temperature and heart rate) and work on doctor to patient communication and implement some emergency feature. | Completed integrating the heart rate sensor. The ADC collects the raw heart rate and runs the data through a heart rate algorithm. The algorithm will measure critical levels. |
5 | 5/12/2015 | Improve app and work on data analysis on the app's side to make the data look meaningful and interesting. | Completed heart rate algorithm and implemented DSP to get a cleaner heart rate. Completed processing accelerometer data to determine the footsteps and orientation. |
6 | 5/19/2015 | Improve and implement the android app and test the entire system from end to end. | Planned |
7 | 5/22/2015 | Demo! | Planned |
Parts List & Cost
Item# | Part | Vendor | Vendor P/N | Cost | Link |
---|---|---|---|---|---|
1 | SJ One Board | Preet | n/a | 79.99$ | http://www.socialledge.com/sjsu/index.php?title=SJ_One_Board |
2 | Pulse Sensor | Sparkfun | SEN-11574 | 24.95$ | https://www.sparkfun.com/products/11574 |
3 | Electric Imp | Sparkfun | WRL-11395 | 29.95$ | https://www.sparkfun.com/products/11395 |
4 | Electric Imp Breakout Board | Sparkfun | BoB-12886 | 12.95$ | https://www.sparkfun.com/products/12886 |
5 | Temperature Sensor-TMP36 | Sparkfun | SEN-10988 | 01.50$ | https://www.sparkfun.com/products/10988 |
- Total Cost: 153.79$
Design & Implementation
Hardware Design
Discuss your hardware design here. Show detailed schematics, and the interface here.
Main Hardware Required
- SJ One Board: Custom made for CMPE 244 class applications.
- Pulse Sensor: Uses plethysmography to get the pulse reading
- MMA8452: Acceleration sensor to get the 3-Axes acceleration
- Electric Imp: Wifi module used to send on-board data to the server
- TMP36: Temperature sensor used to get surface temperature
Hardware Interface
In this section, you can describe how your hardware communicates, such as which BUSes used. You can discuss your driver implementation here, such that the Software Design section is isolated to talk about high level workings rather than inner working of your project.
- Pulse sensor:
- Sends the raw photoplethysmograph signal to the on-board ADC and the signal is sampled every 2ms.
- Accelerometer sensor:
- On-board accelerometer sensor is interfaced via the I2C bus to get the 3-Axes acceleration.
- Temperature sensor:
- External temperature sensor, TMP36, is interfaced via the analog port to get on-body temperature readings.
- Electric Imp:
- Collected data is sent to the server via Electric Imp. UART bus is used for communication between the board and Electric Imp.
Software Design
Show your software design. For example, if you are designing an MP3 Player, show the tasks that you are using, and what they are doing at a high level. Do not show the details of the code. For example, do not show exact code, but you may show psuedocode and fragments of code. Keep in mind that you are showing DESIGN of your software, not the inner workings of it.
- Firmware of the SJ One Board is designed using FreeRTOS. C++ language is used to write the firmware. The drivers provided by the SJ One Board manufacturer are used for low level application. Each task is assigned to collect a measurement, process an algorithm on it and compute the result. To maintain the high accuracy sampling frequency of the pulse sensor, the RIT- repetitive timers, interrupt service routine and binary semaphore are used. To maintain the sampling frequency of the accelerometer and the temperature sensor vTaskDelay is used. Since the clock cycles spent on the tasks are low, using the vTaskDelay was accurate enough to get a reasonably constant sampling frequency of the accelerometer and the temperature sensors.
ALGORITHMS:
1. Heart Rate calculation:
- The heart rate (bpm) is calculated by sensing the photoplethysmograph signal from the Pulse Sensor. By using the adc0_get_reading() function, we are able to read the analog signal from the sensor. Since the analog signal we were getting was noisy, we used a standard median filter with window size of 3 to eliminate the noise. By comparing the three consecutive analog value, we are able to find the peak of the signal. By calculating the average time between the 10 consecutive peaks, the inter-beat interval (IBI) is calculated, from which the heart rate is determined.
- Algorithm Pseudocode:
- Find the peak of the signal, time_of_last_peak = current time.
- Find the next peak, time_of_next_peak = current time
- Interval = time_of_last_peak - time_of_next_peak.
- If interval is < ( Interbeat_Beat_Interval*(3/5) ). Ignore it- this helps to eliminate the dicrotic notch.
- Calculate the Interbeat_Beat_Interval which is average of ten intervals
- Heart rate = 1 min/Interbeat_Beat_Interval and loop back to the: find first peak.........
2. Step Count Algorithm:
- The step count is calculated using the on-board accelerometer. Initial threshold for the step count is calculated based on the R&D, as mentioned in the implementation part.
- Pseudocode:
- Initialize the dynamic Threshold ( refer Analyzing the sensor signal section)
- Calculate the energy, energy = sqrt(x*x + y*y + z*z)
- Find the peak and the trough, amp = peak - trough
- if((amp > dynamic threshold*0.5)&&(current_time-last_step_time > 0.4s)) {step_count++} //Calculate the amplitude
- NOTE: 0.4 sec was picked based on the assumption, possible maximum walking speed is 150 steps/min
- Calculate the harmonic average of the past 5 amplitudes
- Update the dynamic threshold = harmonic average and then loop back to energy calculation.
3. Patient Orientation:
- Pseudocode:
- Read the z acceleration
- If z>700 for 10sec, then patient position = lying down
- If z<700 for 10sec, then patient position = standing.
4. Critical Conditions:
- Determine if the heart rate is critical
- Determine if temperature is normal range
- Determine if emergency is requested by the patient
CLOUD AND THE APP COMMUNICATION
The cloud provided by the Electric Imp is used to pass the data to the mobile app. The Electric Imp consists of two components: 1) The device and 2) The agent-the electric imp cloud server. The device is the actual hardware which sends the data to the agent through WiFi and the mobile app will pull the data from the agent (server). Squirrel language is used to write the software for the agent and the device. The android mobile app is created in Java.
- Logic:
- 1). Device:
- Receive data from the SJ ONE Board through UART.
- Send the data to the agent- Example code: agent.send("HR", HR);
- 2). Agent:
- Receive the data from the device
- Store them in the server.
- On http request data is sent back using the following code:response.send(200,format("heart=%d,temprature=%d,poistion=%d,step=%d,emergency=%d",HRval,tempval,poval,stepval,eoval));
- 1). Device:
Implementation
This section includes implementation, but again, not the details, just the high level. For example, you can list the steps it takes to communicate over a sensor, or the steps needed to write a page of memory onto SPI Flash. You can include sub-sections for each of your component implementation.
Analyzing the Sensor Signals
1. Heart Rate:
- Photoplethysmograph signal of the pulse sensor is obtained by reading the analog signal of the pulse sensor, as previously mentioned. This analog signal is read every 2ms and transferred from the SJ One Board to the computer through UART. The data is then logged into the computer for ten minutes using the Terminal v1.9b software (https://goo.gl/yJZCC).
- This data is imported to Matlab and then plotted as shown in the figure below:
- According to the figure the data obtained contained some noise. Hence, different filters were tried and tested to eliminate the noise. Finally the Standard Median Filter with a window size of 3 was found as the optimal filter with low clock cycles.
- This figure below shows the filtered signal, after the noise was removed:
- This filter is implemented on the SJ ONE Board and the filtered signal is fed into the function which computes the heart rate algorithm.
2. Step Count:
- Read the X,Y,Z acceleration of on board sensor and save the data on the SD card using Storage::append("1:myfile.txt",data,28,0);
- By using the setup as shown in the picture, acceleration of x,y,z for 50 steps was collected on the SD card
- Read the data from the SD card and transfer the data to the computer through UART
- Data was loaded in MatLab and step count algorithm was implemented in Matlab first and this algorithm was verified using this data.
- The figure below shows energy vs time plot while walking: energy = (x*x +y*y +z*z)
- This matlab code was ported to the SJ One board.
Firmware Implementation
Android Implementation
- Step 1 - Setup Java Development Kit (JDK)
- Step 2 - Setup Android SDK
- Step 3 - Setup Eclipse/Studio IDE
- Step 4 - Setup Android Development Tools (ADT) Plugin
- Step 5 - Create Android Virtual Device
- Step 6 - Program the application in Java
- Step 7 - Run the APK file on the android device
Testing & Technical Challenges
Describe the challenges of your project. What advise would you give yourself or someone else if your project can be started from scratch again? Make a smooth transition to testing section and described what it took to test your project.
The main modules for the system were tested on different subjects to check for subject to subject consistency. These tests include:====
- STEP COUNT:
- Step count was tested on two subjects, Subject 1 and Subject 2, and in two different modes, fast and normal mode. On each test 100 steps were made. The table below show the results:
Include sub-sections that list out a problem and solution, such as:
Issues Faced
- Issue #1:
- Issue: Initially, the heart rate algorithm was not working as expected. It would read some junk data values and the values did not align with the on body placement of the sensor. To rule out the problem, the adc value of the pulse sensor was logged and plotted. In the plots, it was noticed that the signal had random sudden spikes(noise). As a result, the heart rate did not read the correct bpm values.
- Solution: To remove the noise spikes from the signal, a standard median filter with window size 3 was implemented.
- Issue #2:
- After integrating the electric Imp with the SJ ONE Board, the heart rate values computed were off again. To solve this issue, we logged and plotted the adc value again. When the Electric Imp was powered from the SJ One Board, periodic spikes were found. But when the Electric Imp was unplugged the signal was clean. From this, it was clear that the noise was coming from the Electric Imp.
- Solution: As a temporary solution, the Electric Imp and the SJ One Board were powered up with two different batteries.
- Issue #3:
- The pulse sensor was very sensitive to the environment light, because of which it gave random heart rate readings even when the sensor was off the body.
- Solution : We assume that the sensor will be on the body while in use.
Conclusion
Conclude your project here. You can recap your testing and problems. You should address the "so what" part here to indicate what you ultimately learnt from this project. How has this project increased your knowledge?
Project Video